Technical Specification
Duty Cycle: Intermittent loading and continuous operation( 10s/30s)
Five Working modes
| Function | Modes | Output Power (load) | V-PP | CF |
| Monopolar Cut | Pure cut | 1-300W @800ohm load | 1.4kV | 1.4 |
| Blend | 1-200W@800ohm load | 1.72KV | 1.7 | |
| Monopolar Coagulation | Coag1 | 1-80W@800ohm load | 2.8KV | 5.8 |
| Coag2 | 1-120W@800ohm load | 640KV | 2.1 | |
| Bipolar | Standard coag | 1-50W@200ohm load | 480V | 1.8 |
Input Power
| Input Voltage | 100-240V~±10% |
| Mains line frequency range | 50-60Hz |
| Power Rating | 1100VA |
| Power consumption | ≤1100VA |
| Fuses(two) | 5A/10A |
| Working frequency | 416KHz |
| Monopolar RF leakage current | <150mA |
| Bipolar RF leakage current | <50mA |
Standards and IEC Classifications
| Class I Equipment IEC 60601-1 |
| Type CF Equipment IEC 60601-1 |
| Spill Resistance IEC 60601-2-2 |
| Electromagnetic Compatibility IEC 60601-1-2 and IEC60601-2-2 |
Dimensions and Weight
| Length: 45cm | Width: 36cm |
| Height : 14cm | Weight: 8kg |
Operating Parameters
| Ambient temperature range: 5℃ to 40℃ |
| Relative humidity: ≤80% |
| Atmospheric pressure: 86kPa to 106kPa |
Configuration Sheet
| Main unit 1set |
| Electrosurgical Pencil 5pieces |
| Electrosurgical Pad 10pieces |
| Electrosurgical pad cable 1piece |
| Double Footswitch 1set |
| Bipolar Forceps with cable 1set |
| Tip Clean Sponge 1piece |
| Power cable 1piece |