SM-8000C Multiparameter Patient Monitor

12.1”TFT screen
  • Parameters

Technical Specification

Application Range

Adult/Pediatric/Neonatal (Necllor Spo2&Suntch Nibp) /Surgery/ Operating Room/ICU/CCU (Separated parameter board)

 

Display

12.1”TFT screen

 

Environment

Operating environment:

Temperature: 0 ~ 40 ℃

Humidity: 15% ~ 80%

Pressure: 86.0 kPa ~ 106.0kPa

Transport and Storage environment :

Temperature: -20 ~ 55 ℃

Humidity: 10% ~ 90%

Pressure: 86.0 kPa ~ 106.0kPa

 

Power Requirements

AC: 100 ~ 240V, 50Hz/60Hz

DC: Built-in rechargeable battery

Battery: 11.1V 24wh Li-ion battery

4h operating after full charge

5min operating after low battery alarm

 

Dimension and Weight

Equipment: 305*162*290mm(about) ; 4kg(about)

Package: 380*350*300mm(about) ; 4.5kg(about)

 

Date Storage

Trend diagram/table: 720h

NIBP review: 10000 events

Wave review: 12h

Alarm review: 200 alarm events

Support drug concentration titration analysis

 

ECG

5 Leads: RA, LA, LL, RL, V

Lead mode: Ⅰ,Ⅱ, Ⅲ, AVR, AVL, AVF, V

Gain: × 1.25mm/mV,2.5mm/mV,

5.0mm/mV, 10mm/mV,

20mm/mV (error 5%)

Sweep speed: 6.25mm/s,12.5mm/s,25mm/s,

50mm/s (error≤ ±10%)

Bandwidth: Surgery 1 ~ 25 Hz

Monitor 0.5 ~ 40 Hz

Diagnostic 0.05 ~ 130 Hz

Sensitivity: >200 V P-P

CMRR≥ 112dB

Differential Input Impedance: >5M

ohm Electrode Offset Potential ±300mV

Polarization Voltage ≥±500mV

Input Impedance ≥5 M

Calibration Voltage: 1mV± 5%

 

SpO2

Measurement range: 0 ~ 100%

Resolution: 1%

Precision: 70% ~ 100% ±2%

0%~69%:no definition given

 

Pulse Rate

Measure/alarm range: 20 ~ 300bpm

Resolution: 1bpm

Accuracy ±3bpm

 

NIBP

Method: Oscillometry

Measure mode: Manual, Auto, STAT

Measure interval in auto

mode:1~5/10/15/30/60/90/120/ 180/240/480min

Measure period in STAT mode: 5min

Pulse rate range: 40 ~ 240bpm

Measure & alarm range:

Adult SYS 40 ~ 280mmHg

DIA 10 ~ 220mmHg

MEAN 20 ~ 240mmHg

Pediatric SYS 40 ~ 220mmHg

DIA 10 ~ 160mmHg

MEAN 20 ~ 170mmHg

Neonatal SYS 40 ~ 135mmHg

DIA 10 ~ 100mmHg

MEAN 20 ~ 110mmHg

Static pressure measurement

Range: 0 ~ 300mmHg

Accuracy 3mmHg

Resolution: 1mmHg Accuracy

Max. Mean Error 5mmHg

Max. Standard Deviation: 8mmHg

Overpressure protection:

Adult 300mmHg

Pediatric 240mmHg

Neonatal 150mmHg

 

ST Segment

ST Segment Range: -2.0mV ~ +2.0mV

Accuracy: Ranging -0.8 mV ~ +0.8mV,

measuring error is ±0.02mV or ±10%,

The larger prevails.

No definition for other ranges.

 

RESP

Method: impedance between RA-LL

Resp Impedance range: 0.3 ~ 3

Base Impedance range: 200 ~ 4000

Bandwidth: 0.3 ~ 2.5Hz

Resp rate:Measurement Range

Adult 0 ~ 120BrPM

Gain: × 0.25, × 0.5, × 1, × 2, × 4

Neonatal/Pediatric 0 ~ 150BrPM

Resolution: 1BrPM

Accuracy: 7 ~ 150BrPM ( 2BrPM)

0 ~ 6BrPM (unspecified)

Apnea Alarm: 10 ~ 40s

 

TEMP

Channel 2

Measure & alarm range: 0 ~ 50 ℃

Resolution: 0.1 ℃

Accuracy (no sensor) ±0.1 ℃ (0~50 ℃)

(excluding sensor error)

 

Standard accessories

ECG cable & electrodes

SpO2 sensor

TEMP probe

NIBP cuff & tube

Power cord

Lithium-ion battery

User manual

 

Optional

Etco2 (Optional)

Dual IBP (Optional)

Nellcor Spo2 (Optional)

Suntch Nibp (Optional)

Recorder (Optional)

Touch Screen (Optional)

Related Products

SM-200V Patient Monitor

12.1” color TFT LCD screen( touchscreen is optional ) 8 waveform display,up to 12-lead ECG analysis Powerful calculation(Hemodynamic, Dose, Oxygenation, Ventilation) Pacemaker detection ST & arrhythmia analysis(17 types) SpO2 support PI, low perfusion 0.2% Night mode, standby mode, venipucture mode Various mounting solutions Wired/Wireless CMS, support HL7 protocol to HIS SpO2 pulse-tone modulation (Pitch Tone) MEWS(Modified Early Warning Score) Graphical & tabular trend review(480 hours), USB data output Rechargeable Lithium-Ion Battery 48 Hours full disclosure waveforms review & print for each patient