Technical Specification
Application Range
It is suitable on adult children and newborn patients .It can be used during the operation, post-anaesthetic recovery after the operation and so on.
Display
7-inch color LCD display
Environment
working environment conditions
Ambient temperature range: +5ºC~40 ºC;
Relative humidity range: 15%~80%, no condensation
Atmospheric pressure range: 60~107kPa;
Shipping and storage conditions
Ambient temperature: -20ºC~+55ºC
Atmospheric pressure: 16kPa~107kPa
Humidity range: 10%~95%
Electrical Specifications
Input: AC 100~240V AC,
Power frequency: 50/60Hz,
Maximum input power: 70VA
Fuse specification: 2AL 16V
Working method: continuous work
Internal lithium battery: d.c 7.4V Not less than 2000 mAh
Dimension and Weight
Product Size 246.5*146.4*50.7mm
Product weight 950g
ECG
Gain: X0.25, X0.5, X1, X2, Auto
Timebaseline selection:
12.5mm/s, 25mm/s, 50mm/s
Sensitivity>200uV (peak-to-peak)
bandwidth:
Diagnostic mode: 0.05~130Hz
Monitoring mode: 0.5~40Hz
Operation mode: 1~20Hz
Common Mode Rejection Ratio:
Diagnostic mode: >90dB
Custody mode: >100dB
Operation mode: >100dB
NIBP
Adult systolic blood pressure: 40~270mmHg
Diastolic blood pressure: 10~210mmHg
Average pressure: 20~230mmHg
Children systolic blood pressure: 40~200mmHg
Diastolic blood pressure: 10~150mmHg
Average pressure: 20~165mmHg
Neonatal systolic blood pressure: 40~135mmHg
Diastolic blood pressure: 10~95mmHg
Average pressure: 20~110mmHg
Static pressure range: 0~300mmHg
Static pressure accuracy : 3mmHg
Blood pressure accuracy:
maximum average error 5mmHg;
maximum standard deviation 8mmHg Overvoltage Protection:
Adult mode: 300 mmHg±10mmHg
Child mode : 240mmHg±10mmHg
Neonatal mode : 150mmHg±10mmHg
SpO2
Measuring range 0~100% Resolution 1%
The accuracy is within the range of 70% to 100%, and the measurement error should be ±4%;
In the range of 0% to 69%, the measurement error is undefined
Pulse Rate
Measuring range 25~250bpm
Resolution 1bpm
Accuracy ±3bpm
Refresh time 1s
Heart rate
15 to 300 bpm for adults (beat / min)
Newborn / child 15-350 bpm (beat / min)
Accuracy: ± 1% or ± 1 bpm, take the large on
Temperature
Range 0~50ºC
Resolution 0.1ºC
Accuracy ±0.1ºC (without probe error)
RESP
Adjustment range for RESP alarm upper limit and lower limit:
Limit and lo
Max. upper limit 120
Min. lower limit 0
single time adjustment 1
Standard accessories
ECG lead line
Cuff
SpO 2 sensor
TEMP probe
Lithium-ion battery
Power cable
Operator’s manual
Optional accessories
Etco2 (optional)
Wired Wifi (optional)