SM-B806 Concentration Central Delivery System

1. power supply: AC220V±10%, frequency: 50Hz±2%, power: 6KW.
2. Water requirement: temperature 10℃~30℃, water quality meets or better than the requirements of YY0572-2015 “water for Hemodialysis and Relate Treatment.
3. Environment: Ambient temperature is 5℃~40℃, relative humidity is not greater than 80%, atmospheric pressure is 700 hPa~1060 hPa, no volatile gas such as strong acid and alkali, no dust and electromagnetic interference, avoid direct sunlight, and ensure good air mobility.
4. Drainage: drainage outlet ≥1.5 inches, the ground needs to do a good job of waterproof and floor drain.
5. Installation: installation area and weight :≥8(width x length =2x4) square meters, the total weight of the equipment loaded with liquid is about 1 ton.
  • Parameters

Technical parameters

1. Preparation of concentrated liquid:automatic water inlet, water inlet error ≤1%;

2. The preparation solution A and B is independent of each other, and consists of liquid mixing tank and storage with transportation respective. The mixing and supplying parts do not interfere each other.

3. The preparation of concentrated solution is fully controlled by PLC, with 10.1 inch full-color touch screen and simple operation interface, which is convenient for medical staff operating;

4. Automatic mixing procedure, working modes such as water injection, timing mixing, perfusion; Fully dissolve A and B powder, and prevent the loss of bicarbonate caused by excessive stirring of B liquid;

5. Filter: filter the undissolved particles in the dialysis solution, make the dialysis solution meet the requirement of hemodialysis, effective ensure the quality of concentrated solution;

6. Fully automatic flushing and one-button disinfection procedures, effectively prevent the breeding of bacteria;

7. Opened disinfectant, the residual of the concentration after the disinfection meet the standard requirements;

8. All valve parts are made of corrosion-resistant materials, which can be soaked for a long time by strong corrosive liquid and have long service life.

9. The product materials meets the requirements of medical and corrosion resistance.

10. Multiple safety protection :ion concentration monitoring, endotoxin filter , stable pressure control, to ensure the safety of patients and dialysis equipment;

11. Mixing according to the actual needing, reduce errors and pollution.

Related Products

SMT-050DH Double column Digital Radiography System (Hand-driven)

Double column Digital Radiography System adopts large span independent ball tube frame design, easy installation, small occupation area, convenient maintenance. Four-way floating bed surface, easy to use and positioning. The lifting and lifting of X-ray source and bidirectional follow-up of horizontal position of bed column can reduce the positioning time and meet the requirements of large-flow clinical examination. Applicable to orthopedics, emergency room, operating room, etc.,As human head, limbs, chest, spine, lumbar spine for medical diagnosis

VSM-3200T 32kW 400mA Veterinary Digital Radiography System

1. Sophisticated technology, high rotating speed, good heat dissipation, high heat capacity, suitable for long-time clinical examination, long product life. 2. High frequency generator with high X-ray inverse frequency, large power and good X-ray quality. 3. Intelligence high voltage control system with various APR. 4. New generation a-Si Csl flat panel detector with high conversion rate achieves digital image acquisition with low dose. 5. High standard digital image processing system provides more post-processing and meets clinical needs.

SM-800L 2-8°C Medicine Refrigerator

Meet the requirements of the new version GSP for medicine storage. Can be used for medicine, reagents and various items that need cold storage. Suitable for hospitals, clinics, blood stations of health and epidemic prevention systems, university laboratories, etc.

SMH-60B 5-part Auto Hematology Analyzer

1. Throughput 60T/H.       2. 14-inch touch screen. 3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting. 4. 3D holographic scattergram displays the accurate 5-part differentiation of WBC. 5. Large storage capacity: 300,000 results(including histogram, scattergram, patient information).

Leave A Message


    Products
    Contact
    WhatsApp
    Email