VSM-HTY23L Ethylene Oxide Sterilizer

HTY series ethylene oxide sterilizer is a special equipment for low-temperature fumigation and sterilization of items enclosed in a sterilization room using ethylene oxide gas at a certain temperature (low temperature), pressure, humidity and a certain time of action. Ethylene oxide gas is a very active chemical disinfectant. It not only achieves broad-spectrum sterilization, but also can achieve reliable sterilization.
  • Parameters

Sterilization principle:

The ethylene oxide sterilizer produces a very specific alkylation effect through the contact between ethylene oxide and microorganisms, thereby inhibiting the normal reaction of microbial reactive groups, causing metabolic disorders and death.

 

Consumables for ethylene oxide sterilizer:

Ethylene oxide gas, ethylene oxide biological indicator, ethylene oxide indicator tape, ethylene oxide indicator inner card, ethylene oxide indicator outer card, ethylene oxide packaging bag.

 

Ethylene oxide tail gas processor:

The ethylene oxide tail gas processor decomposes the residual ethylene oxide tail gas after sterilization into non-toxic carbon dioxide and water vapor through high-temperature catalytic decomposition, which is directly discharged outdoors without the need for high-altitude emission pipelines. The decomposition efficiency is over 99.99%, greatly reducing the emission of ethylene oxide.

 

Technical specification

Serial number
Name
Technical parameter
1 Product
Model:
VSM-HTY23L
2 Power: 1.5 KVA
3 Power
supply:
220V/50HZ
4 Internal size: Φ250*450mm (allowable deviation ± 5%)
5 Dimensions: 700*530*425mm (allowable deviation ± 5%)
6 Volume: 23L
7 Weight: 40kg
8 Display
screen:
4.3-inch color touch screen is used. Comply with
EN61000-6-2:2005, EN61000-6-4:2007 standards; comply with
RoHS lightning surge  ± 1KV, group pulse+2kv electrostatic
contact 4KV, air discharge 8kV. With lightning surge
protection function. Allowable power loss <5ms.
9 Number of
doors and
switching
method:
Single-door leaf-shaped door switch method is used. The
flatness is high, the manual locking is tight, and the sealing
strip of the door ring can be perfectly compacted and sealed
to ensure the sealing during vacuuming. The external
protective door adopts galvanized sheet and
high-temperature plastic powder curing process, and the
upper end of the door is equipped with a switch door sensor
device to increase operating safety.
10 Heating
method:
Annular heating is used to ensure the temperature uniformity
of the sterilization chamber
11 Cavity
structure
and material:
The cavity structure is circular, and the cavity material is made
of high-quality stainless steel with a thickness of ≥2.2mm. It
has superior thermal conductivity and improves the service life
and deformation resistance of the liner. The inner liner adopts
mirror treatment technology, which is beautiful and seamless,
and has a long service life
12 The outer
box material:
Galvanized steel plate and sheet metal paint are used, which is
simple and easy to take care of, and it is not easy to shed paint
and oxidize.
13 The resolution of the humidity sensor is 0.1%
14 The main
body of
thermal
insulation:
The thermal insulation cotton is made of aluminum silicate
cotton, which has the characteristics of low thermal
conductivity, good fire resistance, stable thermal insulation
performance, easy maintenance, disassembly and assembly,
strong anti-aging ability, non-toxic and environmentally
friendly, high-grade appearance and soft texture.
15 Working
temperature:
30℃-60℃
16 working
pressure:
0kpa~ -75KPa
17 working
humidity:
40%~80%
18 Sterilization
time:
3h‐99h
19 residual
discharge
times:
9-99 times Arbitrarily set
20 Display
content:
Temperature, pressure, time, current stage, running time
andAlarm information and the operation of the internal
hardware of the device. You can intuitively feel that those parts
of the current equipment are working again. That is, it isconvenient for equipment maintenance and can intuitively
learn the operation process and principle of the entire
equipment. And provide photos of the actual display interface.
21 Printing
record
content:
Computer printing can record the sterilization time, display
the pressure, temperature, humidity, and sterilization process,
and provide the printing function of the sterilization physical
graph. The application of large-capacity storage to save the
original data, and the printed records can be stored for more
than five years.
22 Sterilization
monitoring
and alarm
information
storage
system:
It has a sterilization record and alarm record traceability
system, and the information can be displayed directly on the
touch screen for relevant personnel to view at any time. Can
save a complete sterilization record of not less than 8000 pot
times
23 sensors: built-in temperature and humidity, pressure sensors,
automatic detection of the corresponding temperature,
humidity and pressure values in the sterilization room.
24 Vacuum

pump:

oil-free vacuum pump with extremely high vacuum degree
and corrosion resistance is used, the evacuation rate reaches
120L/min, and the operating sound is less than 60dB
25 Pipeline
material:
304 stainless steel sanitary pipeline and sanitary clamp
connection
26 sterilizing
agent and
dosing
method:
Built-in gas source: The sterilizing agent is a single-use, 100%
pure ethylene oxide aerosol canister. The dosing method uses
an internal, automatic puncture technology within the
sterilization chamber, ensuring high safety, convenience, and stability.
External gas source: Uses the dedicated connecting pipe
provided with the equipment, connecting one end to the machine and the other end to the gas cylinder. Simply unscrew the knob when adding sterilizing agent.
Note: Both machine models are available for purchase.
27 The safety of the use of the device is greatly improved.
28 Fault
diagnosis:
it has automatic fault detection, fault code content display
alarm, fault sound alarm and fault printing system. The device
has a power-off memory function.
29 languages: The display interface in Chinese and English is used, and other
display languages can be customized. The entire language
system is unified.
30 Traceability
system:
supports docking with the hospital traceability system and
supports a variety of traceability protocols;
31 Password
protection:
Password protection is used to prevent incorrect use by
non-operators. Some of the main process alarms require a
password to unlock the alarm information. After the
sterilization is completed, you also need to enter the password
to open the door.
32 Manual
testing:
it has a manual testing function to facilitate debugging and
maintenance.
33 Leakage
test:
Independent leakage test procedures are available, and
independent leakage tests can be carried out outside the
operation of the equipment. Independent procedures can be
used to detect the equipment.

 

Related Products

SMT-050DH Double column Digital Radiography System (Hand-driven)

Double column Digital Radiography System adopts large span independent ball tube frame design, easy installation, small occupation area, convenient maintenance. Four-way floating bed surface, easy to use and positioning. The lifting and lifting of X-ray source and bidirectional follow-up of horizontal position of bed column can reduce the positioning time and meet the requirements of large-flow clinical examination. Applicable to orthopedics, emergency room, operating room, etc.,As human head, limbs, chest, spine, lumbar spine for medical diagnosis

VSM-3200T 32kW 400mA Veterinary Digital Radiography System

1. Sophisticated technology, high rotating speed, good heat dissipation, high heat capacity, suitable for long-time clinical examination, long product life. 2. High frequency generator with high X-ray inverse frequency, large power and good X-ray quality. 3. Intelligence high voltage control system with various APR. 4. New generation a-Si Csl flat panel detector with high conversion rate achieves digital image acquisition with low dose. 5. High standard digital image processing system provides more post-processing and meets clinical needs.

SM-800L 2-8°C Medicine Refrigerator

Meet the requirements of the new version GSP for medicine storage. Can be used for medicine, reagents and various items that need cold storage. Suitable for hospitals, clinics, blood stations of health and epidemic prevention systems, university laboratories, etc.

SMH-60B 5-part Auto Hematology Analyzer

1. Throughput 60T/H.       2. 14-inch touch screen. 3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting. 4. 3D holographic scattergram displays the accurate 5-part differentiation of WBC. 5. Large storage capacity: 300,000 results(including histogram, scattergram, patient information).

Leave A Message


    Products
    Contact
    WhatsApp
    Email