SM-5W Oxygen Concentrator

Recommended maximum oxygen flow: 5L/min.
Flow scope when the nominal pressure at the outlet is 70kPa: 0.5L/min~5L/min.
Apply 70kPa back pressure at the recommended maximum flow, the flow reaches accuracy of ±10% to the indicated flow.
Oxygen concentration when the nominal pressure at the output is zero (achieve the required oxygen concentration level within 30 minutes of the initial startup): the oxygen yield is 0.5L/min to 5L/min, the oxygen concentration is ≥90%.
Output pressure: 0.02MPa~0.07MPa.
Release pressure of compressor’s safety valve: 280kPa±50kPa.
Noise: ≤ 43dB(A)
Maximum nebulization rate: ≥ 0.1mL / min (for model with nebulizer).
Power supply voltage: ~110V/220V, 50/60HZ
Input power: 450VA
Net weight: 16.5 kg
Dimensions: 39×25×47.6(cm)
  • Parameters

Main technical specifications

1. Recommended maximum oxygen flow: 5L/min.

2. Flow scope when the nominal pressure at the outlet is 70kPa: 0.5L/min~5L/min.

3. Apply 70kPa back pressure at the recommended maximum flow, the flow reaches accuracy of ±10%

to the indicated flow.

4. Oxygen concentration when the nominal pressure at the output is zero (achieve the required oxygen concentration level within 30 minutes of the initial startup): the oxygen yield is 0.5L/min to 5L/min, the oxygen concentration is ≥90%.

5. Output pressure: 0.02MPa~0.07MPa

6. See Figure 1 for function table of oxygen concentration and oxygen flow when the nominal pressure at the output is zero.

7. Release pressure of compressor’s safety valve: 280kPa±50kPa

8. Noise: ≤ 43dB(A)

9. Maximum nebulization rate: ≥ 0.1mL / min (for model with nebulizer)

10. Power supply voltage: ~110V/220V, 50/60HZ

11. Input power: 450VA

12. Net weight: 16.5 kg

13. Dimensions: 39×25×47.6(cm)

14. Altitude: the oxygen concentration will not be reduced when it is from sea level to 1828m, but the oxygen concentration will be less than 90% when it is from 1828m to 4000m

15. Safety system:

▲ Shutdown due to current overload or loose connections

▲ Shutdown due to high temperature of compressor;

▲ Low oxygen concentration alarm;

16. Shortest working hours: not less than 30 minutes.

17. Electrical classification: Class II equipment, type B application part

18. Running mode: continuous operation

19. Normal working environment

▲ Ambient temperature range: 10 ℃ to 40 ℃;

▲ Relative humidity range: ≤ 80%;

▲ Atmospheric pressure range: 860hPa to 1060hPa.

Notice: When the storage and transportation temperature is below 10℃, the oxygen concentrator shall be placed in normal working temperature for more than 4 hours before operation.

20. Oxygen outlet temperature: ≤ 46℃

21. Suggestion: the length of the oxygen tube shall not exceed 6 meters, and the oxygen tube can not be bent or deflated.

22. Storage and transportation environment:

▲ Ambient temperature range: -40℃~+70 ℃;

▲ Relative humidity range: 10% to 100%, and no condensation;

▲ Atmospheric pressure range: 860hPa to 1060hPa.

Notice: The oxygen concentrator shall be stored inside the room where there is no strong sunlight, no corrosive gases but well ventilated. Do not shake or place upside down the oxygen concentrator during transportation.

 

List of packaging

1. oxygen concentrator 1 piece

2. disposable nasal oxygen tube 2 pieces

3. air filter 6 pieces

4. remote control 1 piece

5. O-shaped sealing ring of humidifier 1 piece

6. Humidifier plug 1 piece

7. Operation Instruction 1 copy

8. Primary filter (optional) 2 pieces

9. Fire damper 1 set

The oxygen concentrator with nebulizer is additionally provided with air pipe, nebulization medicine cup, nozzle or face mask.

Related Products

CN-02A Handheld Nebulizer

Compact and portable, low noise design. Atomized particles can reach deep into the lungs. High sensitivity water detection, small liquid residue is not wasted. Intelligent low power mode with long standby time. Self-containd cleaning mode.

CN-02 Air Compressed Nebulizer

Lightweight and portable, atomize anytime, anywhere. 60dB soft sound atomization, multiple noise reduction, soft sound. Adjustable atomization cup, the amount of fog can bead-justed as you like. The atomized particles are fine and ultra-fine atomized, reaching deep into the lungs. Mask/mouthpiece, two modes selectable. Medical grade pp material, can directly contact with the baby's skin. The atomization rate and spray volume are constant, gentle and not throat-choking.