Technical Specification
| Overall Performance |
Throughput | 400 tests/hour, ISE(Optional) |
| Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry | |
| System | Optional open or close system, continuous working for 24 hours | |
| Sample&Reagent System | Sample Positions | 49 sample positions |
| Reagent Position | 56 reagent positions | |
| Sample Volume | 2~70μl | |
| Reagent Volume | 20~350μl | |
| Probe | With anti-collision function, liquid level detection function | |
| Probe Washing | Automatic washing interior and exterior | |
| Reagent Cooling | 2~12℃ for 24 hours | |
| Reaction System | Temperature Control | 37±0.2℃, real-time monitoring |
| Cuvettes | 120 reusable cuvettes | |
| Mixer Probe | Independent stirring, automatic frequency conversion | |
| Washing | Automatic cuvettes washing | |
| Optical System | Light Source | Halogen lamp, water-cooling |
| Spectrophotometry | Post-spectral spectrophotometry | |
| Wavelength | 340~800nm, <±1nm | |
| Absorbance | 0~3.5Abs | |
| Calibration&QC | Calibration | Linear: K factor, 1-point, 2-point and multipoint linear
Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P |
| Quality Control | Real-time QC, Westgard multi rule, Cumulative sum check, Twin plot(2D) | |
| Data Management | Software | Windows 7/8/10, 32 or 64 bit |
| LIS System | Bi-direction, support HL7 protocol | |
| Interface | LAN port access | |
| Printer | External, multiple reporting mode available | |
| Working Conditions | Power Supply | AC220V±10%, 50/60Hz(Standard)
110V±10%, 60Hz(Optional), 650VA |
| Temperature | 15~30℃(±2℃) | |
| Water Consumption | Deionized water<10L/h | |
| Humidity | 40~85% | |
| Size&Weight | External Size(W*D*H) | 973*645*520mm |
| Net Weight | 96kg | |
| Package Size(W*D*H) | 1130*735*1040mm | |
| Gross Weight | 141kg |