Technical Specification
| Item | Descriptions | ||||
| General | Throughput | 60 samples/hour | |||
| Principle | Tri-angle laser scattering and sheath flow for WBC differentiation andc counting, impedance for RBC and PLT, cyanide-free method for HGB | ||||
| Channels | 2 | ||||
| Aperture diameter | RBC/PLT chamber: 70um | ||||
| Parameters | 25+4 parameters: WBC, Lym%, Mon%, Neu%, Bas%, Eos%, Lym#,Mon#, Neu#, Eos#, Bas#, RBC, HGB, | ||||
| HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC, LIC%, LIC#, | |||||
| ALY%, ALY# | |||||
| Histograms | 4 scatter-grams, 2 histograms(including RBC, PLT histograms) | ||||
| HGB lamp | LED530nm | ||||
| Sample volume | Whole blood mode: 20ul, Pre-diluted mode: 20ul | ||||
| Software | Calibration | Manual, auto and fresh blood calibration | |||
| Counting modes | Whole blood, capillary whole blood, prediluted | ||||
| Auto print, manual print | |||||
| LIS | Support LIS | ||||
| Reagent | Reagent | Standard: Diluent: 20L, DIFF Lyse: 500ml, LH Lyse: 100mL | |||
| Cleanser | Standard: Probe cleanser: 50ml | ||||
| Specifications | Input | 10.4 inch touch screen, keyboard, mouse | |||
| Output | Internal printer, support external printer | ||||
| Printer paper | 57×35mm | ||||
| Interface | RS232 port, 4 USB ports, 1 network ports | ||||
| Storage | 50,000 results with histogram | ||||
| Dimension | 475 x 325 x 465mm | ||||
| Blockage clear | High voltage, high pressure flush | ||||
| Net weight | 28KG | ||||
| Others | Work temperature | 10-30°C | |||
| Power requirement | 100~240 AC, 50/60Hz, 200VA | ||||
| Relative humidity | ≤85﹪ | ||||
| Atmospheric pressure | 70.0kPa-106.0kPa | ||||
| Performance | Carry-over rate | Parameter | CV | ||
| WBC | ≤ 0.5% | ||||
| RBC | ≤ 0.5% | ||||
| HGB | ≤ 0.6% | ||||
| PLT | ≤ 1.0% | ||||
| Background | Parameter | Background | |||
| WBC | ≤ 0.2×109/L | ||||
| RBC | ≤ 0.02×1012/L | ||||
| HGB | ≤ 1 g/L | ||||
| HCT | ≤ 0.5% | ||||
| PLT | ≤10×109/L | ||||
| Precision | Parameter | Whole blood sample CV (CV/absolute deviation d) | Pre-diluted sample CV (CV/absolute deviation d) | Measurement range | |
| WBC | ≤2.0% | ≤4.0% | (4.00~15.00)*109/L | ||
| RBC | ≤1.5% | ≤2.0% | (3.50~6.00)*1012/L | ||
| HGB | ≤1.5% | ≤2.0% | (110~180)g/L | ||
| PLT | ≤4.0% | ≤8.0% | (150~500)*109/L | ||
| Linearity | Parameter | Linearity Range | Linearity(Whole blood) | ||
| WBC | (0.00~100.00)*109/L | ±5% | |||
| RBC | (0.00~8.00)*1012/L | ±5% | |||
| HGB | (0~250)g/L | ±2% | |||
| PLT | (0~1000)*109/L | ±8% | |||