Introduction
The 5-part hematology analyzer is an in vitro diagnostic device developed based on laser flow cytometry analysis, nucleic acid fluorescence staining and other technologies, which can quantitatively and qualitatively analyze components such as red blood cells, white blood cells and platelets in the blood.
Technical Specification
| Throughput | 60 Tests/hour | |||
| Assay Items | 5 parts, 29 parameters, 3 histograms, 3D scattergram | |||
| Principle | Tri-angle laser scatter, Flow cytometry method, 3D scattergram analysis, Impedance method for RBC and PLT counting, Cyanide-free method for HGB test | |||
| Test Mode | CBC mode, CBC+DIFF mode
Venous whole blood, Capillary whole blood and Prediluted |
|||
| Sample Volume | CBC+DIFF mode: ≤20μl
CBC mode: ≤10μl |
|||
| Parameters | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PCT, PDW, P-LCR, P-LCC, NEU%, LYM%, MON%, EOS%, BAS%, NEU#, LYM#, MON#, EOS#, BAS#
4 research parameters: ALY%, ALY#, LIC%, LIC# |
|||
| Performance | Item | Linearity Range | Carry Over | CV |
| WBC | 0~300*109/L | ≤0.5% | ≤2.0% | |
| RBC | 0~8.6*1012/L | ≤0.5% | ≤1.5% | |
| HGB | 0~260g/L | ≤0.5% | ≤1.5% | |
| PLT | 0~3000*109/L | ≤1.0% | ≤4.0% | |
| Storage | 300,000 results including histogram, scattergram and patient information | |||
| Interface | 4 USB ports, 1 LAN port Bi-direction LIS, support HL7 protocol, Internal RFID reader | |||
| Power Supply | AC100-240V,50/60Hz, 300VA | |||
| Package Size(W*D*H) | 575*490*690mm | |||
| Gross Weight | 46kg | |||
| Reagent Package Size | 330*330*330m, 400*290*240mm, 400*290*240mm | |||
| Reagent Gross Weight | 22kg, 2kg, 2kg | |||