Technical Specification
| Single-channel Adjustable Volume Pipettes | ||||||
| Volume Range | Increment | Test Volume | Maximum permissible systematic error
(Inaccuracy) |
Maximum permissible random error
(Imprecision) |
||
| μL | μL | μL | % | μL | % | μL |
|
0.1-2.5 |
0.05 |
2.5 | ±2.50 | ±0.0625 | ±2.00 | ±0.05 |
| 1.25 | ±3.00 | ±0.0375 | ±3.00 | ±0.0375 | ||
| 0.25 | ±12.00 | ±0.03 | ±6.00 | ±0.015 | ||
|
0.5-10 |
0.1 |
10 | ±1.00 | ±0.1 | ±0.80 | ±0.08 |
| 5 | ±1.50 | ±0.075 | ±1.50 | ±0.075 | ||
| 1 | ±2.50 | ±0.025 | ±1.50 | ±0.015 | ||
|
2-20 |
0.5 |
20 | ±0.90 | ±0.18 | ±0.04 | ±0.08 |
| 10 | ±1.20 | ±0.12 | ±1.00 | ±0.1 | ||
| 2 | ±3.00 | ±0.06 | ±2.00 | ±0.04 | ||
|
5-50 |
0.5 |
50 | ±0.60 | ±0.3 | ±0.30 | ±0.15 |
| 25 | ±0.90 | ±0.225 | ±0.60 | ±0.15 | ||
| 5 | ±2.00 | ±0.1 | ±2.00 | ±0.1 | ||
|
10-100 |
1 |
100 | ±0.80 | ±0.8 | ±0.15 | ±0.15 |
| 50 | ±1.00 | ±0.5 | ±0.40 | ±0.2 | ||
| 10 | ±3.00 | ±0.3 | ±1.50 | ±0.15 | ||
|
20-200 |
1 |
200 | ±0.60 | ±1.2 | ±0.15 | ±0.3 |
| 100 | ±0.80 | ±0.8 | ±0.30 | ±0.3 | ||
| 20 | ±3.00 | ±0.6 | ±1.00 | ±0.2 | ||
|
50-200 |
1 |
200 | ±0.60 | ±1.2 | ±0.15 | ±0.3 |
| 100 | ±0.80 | ±0.8 | ±0.30 | ±0.3 | ||
| 50 | ±1.00 | ±0.5 | ±0.40 | ±0.2 | ||
|
100-1000 |
5 |
1000 | ±0.60 | ±6 | ±0.20 | ±2 |
| 500 | ±0.70 | ±3.5 | ±0.25 | ±1.25 | ||
| 100 | ±2.00 | ±2 | ±0.70 | ±0.7 | ||
|
200-1000 |
5 |
1000 | ±0.60 | ±6 | ±0.20 | ±2 |
| 500 | ±0.70 | ±3.5 | ±0.25 | ±1.25 | ||
| 200 | ±0.90 | ±1.8 | ±0.30 | ±0.6 | ||
|
500-5000 |
50 |
5000 | ±0.50 | ±25 | ±0.15 | ±7.5 |
| 2500 | ±0.60 | ±15 | ±0.30 | ±7.5 | ||
| 500 | ±0.70 | ±3.5 | ±0.30 | ±1.5 | ||
|
1000-5000 |
50 |
5000 | ±0.50 | ±25 | ±0.15 | ±7.5 |
| 2500 | ±0.60 | ±15 | ±0.30 | ±7.5 | ||
| 1000 | ±0.70 | ±7 | ±0.30 | ±3 | ||
|
1000-10000 |
100 |
10000 | ±0.60 | ±60 | ±0.20 | ±20 |
| 5000 | ±1.20 | ±60 | ±0.30 | ±15 | ||
| 1000 | ±3.00 | ±30 | ±0.60 | ±6 | ||
| 2000-10000
Do not apply to MicroPette plus |
100 | 10000 | ±0.60 | ±60 | ±0.20 | ±20 |
| 5000 | ±1.20 | ±60 | ±0.30 | ±15 | ||
| 2000 | ±3.00 | ±60 | ±0.60 | ±12 | ||
| *2000-20000 | 100 | 2000 | ±6.00 | ±120 | ±3.00 | ±60 |
| 10000 | ±1.20 | ±120 | ±0.60 | ±60 | ||
| 20000 | ±0.60 | ±120 | ±0.30 | ±60 | ||
| 8-channel Adjustable Volume Pipettes | ||||||
| Volume
Range |
IncrementTest Volume |
Maximum permissible
systematic error (Inaccuracy) |
Maximum permissible
random error (Imprecision) |
|||
| % | μL | % | μL | |||
|
0.5-10μL |
0.1μL |
10μL | ±1.50 | ±0.15 | ±1.50 | ±0.15 |
| 5μL | ±2.50 | ±0.125 | ±2.50 | ±0.125 | ||
| 1μL | ±4.00 | ±0.04 | ±4.00 | ±0.04 | ||
|
5-50μL |
0.5μL |
50μL | ±1.00 | ±0.5 | ±0.50 | ±0.25 |
| 25μL | ±1.50 | ±0.375 | ±1.00 | ±0.25 | ||
| 5μL | ±3.00 | ±0.15 | ±2.00 | ±0.1 | ||
|
10-100μL |
1μL |
100μL | ±0.80 | ±0.8 | ±0.30 | ±0.3 |
| 50μL | ±1.00 | ±0.5 | ±0.80 | ±0.4 | ||
| 10μL | ±3.00 | ±0.3 | ±2.00 | ±0.2 | ||
|
50-300μL |
5μL |
300μL | ±0.70 | ±2.1 | ±0.25 | ±0.75 |
| 150μL | ±1.00 | ±1.5 | ±0.50 | ±0.75 | ||
| 50μL | ±1.50 | ±0.75 | ±0.80 | ±0.4 | ||
12-channel Adjustable Volume Pipettes
| 12-channel Adjustable Volume Pipettes | ||||||
| Volume
Range |
Increment Test Volume |
Maximum permissible
systematic error (Inaccuracy) |
Maximum permissible
random error (Imprecision) |
|||
| % | μL | % | μL | |||
|
0.5-10μL |
0.1μL |
10μL | ±1.50 | ±0.15 | ±1.50 | ±0.15 |
| 5μL | ±2.50 | ±0.125 | ±2.50 | ±0.125 | ||
| 1μL | ±4.00 | ±0.04 | ±4.00 | ±0.04 | ||
|
5-50μL |
0.5μL |
50μL | ±1.00 | ±0.5 | ±0.50 | ±0.25 |
| 25μL | ±1.50 | ±0.375 | ±1.00 | ±0.25 | ||
| 5μL | ±3.00 | ±0.15 | ±2.00 | ±0.1 | ||
|
10-100μL |
1μL |
100μL | ±0.80 | ±0.8 | ±0.30 | ±0.3 |
| 50μL | ±1.00 | ±0.5 | ±0.80 | ±0.4 | ||
| 10μL | ±3.00 | ±0.3 | ±2.00 | ±0.2 | ||
|
50-300μL |
5μL |
300μL | ±0.70 | ±2.1 | ±0.25 | ±0.75 |
| 150μL | ±1.00 | ±1.5 | ±0.50 | ±0.75 | ||
| 50μL | ±1.50 | ±0.75 | ±0.80 | ±0.4 | ||
| Fixed Volume Pipettes | ||||||
| Volume
Range |
Increment Test Volume |
Maximum permissible
systematic error (Inaccuracy) |
Maximum permissible
random error (Imprecision) |
|||
| % | μL | % | μL | |||
| 5μL | – | 5μL | ±1.3 | ±0.065 | ±1.2 | ±0.06 |
| 10μL | – | 10μL | ±0.8 | ±0.08 | ±0.8 | ±0.08 |
| 20μL | – | 20μL | ±0.6 | ±0.12 | ±0.5 | ±0.1 |
| 25μL | – | 25μL | ±0.5 | ±0.125 | ±0.3 | ±0.075 |
| 50μL | – | 50μL | ±0.5 | ±0.25 | ±0.3 | ±0.15 |
| 100μL | – | 100μL | ±0.5 | ±0.5 | ±0.3 | ±0.3 |
| 200μL | – | 200μL | ±0.4 | ±0.8 | ±0.2 | ±0.4 |
| 250μL | – | 250μL | ±0.4 | ±1.0 | ±0.2 | ±0.5 |
| 500μL | – | 500μL | ±0.3 | ±1.5 | ±0.2 | ±1.0 |
| 1000μL | – | 1000μL | ±0.3 | ±3.0 | ±0.2 | ±2.0 |
| 2000μL | – | 2000μL | ±0.3 | ±6.0 | ±0.15 | ±3.0 |
| 5000μL | – | 5000μL | ±0.3 | ±15 | ±0.15 | ±7.5 |