Technical Specification
Power supply: AC220V-230V/50Hz or AC110V-120V, 50/60Hz, 75VA
Light source: Blue and yellow LED
Dominant wavelength range of blue light: 450-470nm
Dominant wavelength range of yellow light: 580-595nm
Effective surface area: 50cm×25cm
Life time of LED: 20000hrs
General bilirubin radiation of the effective surface area (Irradiation distance 30cm):
High irradiance mode: ≥54uW/cm²/nm
Low irradiance mode: ≥30uW/cm²/nm
Uniformity of the total irradiance for bilirubin: >0.4
Adjustable range for the height of lamp module: 1350mm-1680mm
Working mode: count up, count down, gear choice mode
Radiation angle of the phototherapy lamp adjustable: 0-90°up from horizontal and can be locked
Built-in timer range: 0~99999.9hrs
Working noise: ≤50dB(A)
Operating request
Temperature range: 18℃~35℃
Relative humidity: 30%~75%RH
Atmospheric Pressure: 800~1060hpa
Standard Configuration
Lamp module and mobile stand
Product Packaging
Package: the whole unit is packed into one carton
Size: 1375mm x 685mm x 230mm
Cross Weight: 23Kg