Technical Specification
Power requirement: AC220-230V/50Hz or AC110-120V, 50/60Hz, 850VA
Auxiliary power output: AC220-230V/50Hz, maximum current allowed: 1.6A
Incubator Module
Air temperature control range: 25.0°C-37.0°C (override mode 37.1°C-39.0°C, optional)
Baby temperature control range: 34.0°C-37.0°C (override mode 37.1°C-38.0°C, optional)
Temperature sensor display range: 5°C-65°C
Warm-up time (with the ambient temperature at + 22°C): No longer than 30 minutes
Accuracy of skin temperature sensor: within ± 0.2°C (optional)
Temperature Variation: ≤0.5℃
Difference between the air temperature at one point and the average air temperature: No greater than 0.8°C
Difference between the average air temperature and the control temperature: No greater than 1.0°C
Temperature uniformity (mattress horizontally placed): No greater than 0.8°C
Temperature uniformity (mattress titled): No greater than 1.0°C
Mattress size: 633mm×342mm
Weight display accuracy:±1% (optional)
Inclination of the bassinet: ±12°
Environmental noise in the compartment in incubator mode:
Under stable temperature ≤ 45dB(A)
Under other conditions ≤ 50dB(A)
Failure alarm: power failure alarm, fan motor alarm, sensor alarm, deviation alarm, over-temperature alarm, system alarm and so on
CO2 concentration in the compartment: Lower than 0.5% when the CO2 mixture is pumped in at the speed of 750 ml/min at 10 cm above the center of the mattress
Air flow speed above the mattress: no greater than 0.35 m/s
Environmental Requirement
Operating range: 20°C-30°C
Relative Humidity: 30%RH-75%RH
Atmospheric pressure range:800hPa-1060hPa
Ambient air flow speed: < 0.3m/s
Product Dimension
Dimension:
L1170mm × W620mm × H1620mm (with stand), L1170mm × W620mm × 1550mm-1700mm (with VHA stand)
Distance from the bassinet to the floor:
980mm
950mm-1100mm
Hood size: L 856mm×400mm×454mm
Wheel size(diameter): 100mm
Weight: 103kg (with stand), 113kg (with VHA stand)
Function Introduction
Two control modes: air mode and baby mode controlled by micro-computer
Set temperature, air temperature and baby temperature are displayed separately
Passive humidification
Inclination if bassinet is adjustable
Independent over-temperature protection system
Multiple failure alarm indications
Independent locking device for the front door
RS-232 connector for data output
Oxygen input interface, Oxygen supply system
Low noise DC motor
Air filter system able to get filtration of most common dust in the air circulation and easy to replace
The pressure difference between the spiral duct and the tributary centrifugal fan ensures that fresh air is always inhaled
Standard Configuration
Main body (including baby compartment, base, controller, I.V. pole and tray), stand
Optional Configuration
>37°C override mode, VHA stand, weighing system