• Parameters

Technical Specification

 

Power requirement: AC220-230V/50Hz or AC110-120V, 50/60Hz, 850VA

Auxiliary power output: AC220-230V/50Hz, maximum current allowed: 1.6A

 

Incubator Module

Air temperature control range: 25.0°C-37.0°C (override mode 37.1°C-39.0°C, optional)

Baby temperature control range: 34.0°C-37.0°C (override mode 37.1°C-38.0°C, optional)

Temperature sensor display range: 5°C-65°C

Warm-up time (with the ambient temperature at + 22°C): No longer than 30 minutes

Accuracy of skin temperature sensor: within ± 0.2°C (optional)

Temperature Variation: ≤0.5℃

Difference between the air temperature at one point and the average air temperature: No greater than 0.8°C

Difference between the average air temperature and the control temperature: No greater than 1.0°C

Temperature uniformity (mattress horizontally placed): No greater than 0.8°C

Temperature uniformity (mattress titled): No greater than 1.0°C

Mattress size: 633mm×342mm

Weight display accuracy:±1% (optional)

Inclination of the bassinet: ±12°

Environmental noise in the compartment in incubator mode:

Under stable temperature ≤ 45dB(A)

Under other conditions ≤ 50dB(A)

Failure alarm: power failure alarm, fan motor alarm, sensor alarm, deviation alarm, over-temperature alarm, system alarm and so on

CO2 concentration in the compartment: Lower than 0.5% when the CO2 mixture is pumped in at the speed of 750 ml/min at 10 cm above the center of the mattress

Air flow speed above the mattress: no greater than 0.35 m/s

 

Environmental Requirement

Operating range: 20°C-30°C

Relative Humidity: 30%RH-75%RH

Atmospheric pressure range:800hPa-1060hPa

Ambient air flow speed: < 0.3m/s

 

Product Dimension

Dimension:

L1170mm × W620mm × H1620mm (with stand), L1170mm × W620mm × 1550mm-1700mm (with VHA stand)

Distance from the bassinet to the floor:

980mm

950mm-1100mm

Hood size: L 856mm×400mm×454mm

Wheel size(diameter): 100mm

Weight: 103kg (with stand), 113kg (with VHA stand)

 

Function Introduction

Two control modes: air mode and baby mode controlled by micro-computer

Set temperature, air temperature and baby temperature are displayed separately

Passive humidification

Inclination if bassinet is adjustable

Independent over-temperature protection system

Multiple failure alarm indications

Independent locking device for the front door

RS-232 connector for data output

Oxygen input interface, Oxygen supply system

Low noise DC motor

Air filter system able to get filtration of most common dust in the air circulation and easy to replace

The pressure difference between the spiral duct and the tributary centrifugal fan ensures that fresh air is always inhaled

 

Standard Configuration

Main body (including baby compartment, base, controller, I.V. pole and tray), stand

 

Optional Configuration

>37°C override mode, VHA stand, weighing system

Related Products

SMT-050DH Double column Digital Radiography System (Hand-driven)

Double column Digital Radiography System adopts large span independent ball tube frame design, easy installation, small occupation area, convenient maintenance. Four-way floating bed surface, easy to use and positioning. The lifting and lifting of X-ray source and bidirectional follow-up of horizontal position of bed column can reduce the positioning time and meet the requirements of large-flow clinical examination. Applicable to orthopedics, emergency room, operating room, etc.,As human head, limbs, chest, spine, lumbar spine for medical diagnosis

VSM-3200T 32kW 400mA Veterinary Digital Radiography System

1. Sophisticated technology, high rotating speed, good heat dissipation, high heat capacity, suitable for long-time clinical examination, long product life. 2. High frequency generator with high X-ray inverse frequency, large power and good X-ray quality. 3. Intelligence high voltage control system with various APR. 4. New generation a-Si Csl flat panel detector with high conversion rate achieves digital image acquisition with low dose. 5. High standard digital image processing system provides more post-processing and meets clinical needs.

SM-800L 2-8°C Medicine Refrigerator

Meet the requirements of the new version GSP for medicine storage. Can be used for medicine, reagents and various items that need cold storage. Suitable for hospitals, clinics, blood stations of health and epidemic prevention systems, university laboratories, etc.

SMH-60B 5-part Auto Hematology Analyzer

1. Throughput 60T/H.       2. 14-inch touch screen. 3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting. 4. 3D holographic scattergram displays the accurate 5-part differentiation of WBC. 5. Large storage capacity: 300,000 results(including histogram, scattergram, patient information).

Leave A Message


    Products
    Contact
    WhatsApp
    Email