SMT-6AP Fetal & Maternal Monitor

  • Parameters

Technical Specifications

FHR:

Ultrasound Frequency: 1MHz±1KHz

Ultrasound input intensity: < 1mW/cm2

Range: 30~240bpm

Accuracy: ±1bpm

Alarm limit & tolerance: Max:(30~240)bpm, tolerance: ±5%.

Min:(30~240)bpm, tolerance: ±5%

 

TOCO

Measurement range: 0~100%

Resolution: 1%

Pressure tolerance: ≤±8%

Zero-mode: Manual

 

Fetal Movement

Count mode: Manual, automatic and automatic fetal movement waveform

Printing paper: Foldable Z-type thermal paper, Width: 112mm

Sweep speed: 1cm/min,2cm/min,3cm/min,25mm/sec, quick recall printing

 

ECG

Input: 3 / 5 wires ECG cable

Lead section: I, II, III, aVR, aVL, aVF, V

Gain selection: ×1/4, ×1/2, ×1, ×2, ×4 and automatic

Sweep speed: 3.125, 6.25, 12.5, 25, 50mm/s

Heart rate range: 15~350bpm

Calibration: ±1mv

Accuracy: ±1bpm or ±1% (select the bigger data)

 

SpO2

Display type: Waveform, Data

Measurement range: 0~100%

Accuracy: ±3% (between 70%~100%)

SpO2 resolution: 1%

Pulse rate range: 25~250bpm

Accuracy: ±1bpm or ±2% (select the bigger data)

PR Resolution: 1bpm

 

PI:

Measurement range: 0.00~20.0%

Accuracy: not defined

Resolution: 0.00~9.99%: 0.01%

10.0~20.0%: 0.1%

 

NIBP

Philosophy: Oscillometric method

Measurement parameter: Systolic, Diastolic, Mean

Measurement method: Manual, Auto, Continuous

Measuring range:

SYS: 30-270mmHg

DIA: 10-200mmHg

MAP:20-220mmHg

Resolution: 1mmHg

Accuracy: Max average error: ±5mmHg

Max standard deviation:8mmHg

 

2-Temperature (Rectal & Surface)

Number of channel: 2 channel

Measurement range: 0°C ~50°C

Accuracy: ±0.2°C

Resolution: 0.1 °C

Display: T1, T2, ΔT

Unit: °C/℉selection

 

2-Respiration (Impedance & Nasal Tube)

Measurement range: 0~120rpm

Accuracy: ±1rpm or ±5%

(select the bigger data)

Resolution: 1rpm

 

Standard configuration:

FHR, TOCO, FM, Built-in printer ECG, NIBP, SPO2, 2RESP, 2TEMP, HR/PR, Rechargeable battery

 

Optional configuration:

Twin FHR, Wireless Probe, Fetal Wake Up, Fisher Auto Fetal-Analysis, Touch Screen, Wireless Network, 12-Volt Power Supply, High Capacity Battery, Wall Mount, Moving Cart,PI

 

POWER REQUIREMENT

Working voltage    100~240VAC, 50/60Hz

Power rate         80VA

Gross Weight:      7.5kg

Carton Size:        43*35*30cm

Display:           10.4” TFT LCD

Resolution:        800*600 pixels

Related Products

SMT-050DH Double column Digital Radiography System (Hand-driven)

Double column Digital Radiography System adopts large span independent ball tube frame design, easy installation, small occupation area, convenient maintenance. Four-way floating bed surface, easy to use and positioning. The lifting and lifting of X-ray source and bidirectional follow-up of horizontal position of bed column can reduce the positioning time and meet the requirements of large-flow clinical examination. Applicable to orthopedics, emergency room, operating room, etc.,As human head, limbs, chest, spine, lumbar spine for medical diagnosis

VSM-3200T 32kW 400mA Veterinary Digital Radiography System

1. Sophisticated technology, high rotating speed, good heat dissipation, high heat capacity, suitable for long-time clinical examination, long product life. 2. High frequency generator with high X-ray inverse frequency, large power and good X-ray quality. 3. Intelligence high voltage control system with various APR. 4. New generation a-Si Csl flat panel detector with high conversion rate achieves digital image acquisition with low dose. 5. High standard digital image processing system provides more post-processing and meets clinical needs.

SM-800L 2-8°C Medicine Refrigerator

Meet the requirements of the new version GSP for medicine storage. Can be used for medicine, reagents and various items that need cold storage. Suitable for hospitals, clinics, blood stations of health and epidemic prevention systems, university laboratories, etc.

SMH-60B 5-part Auto Hematology Analyzer

1. Throughput 60T/H.       2. 14-inch touch screen. 3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting. 4. 3D holographic scattergram displays the accurate 5-part differentiation of WBC. 5. Large storage capacity: 300,000 results(including histogram, scattergram, patient information).

Leave A Message


    Products
    Contact
    WhatsApp
    Email