SMT-350AR Argon Plasma Electrosurgical unit

The 416KHz(+10kHz) high frequency current generated by this argon plasma diathermy machine can performs cut and coagulation for biological tissues.
It integrates cutting, coagulation and AIC argon technology.
It has seven working modes: Cut: Pure cut, Blend 1, Blend 2; Monopolar coagulation: Spray Coagulation, Soft Coagulation, Argon beam coagulation; and Bipolar coagulation.
  • Parameters

Introduction:

The output of the machine is fully suspended. It has two independent and isolated type CF anti-defibrillation applications (monopolar and bipolar), non-AP, non-APG type common equipment, 110V-220V (+10%) power supply.

General Description:

SMT-350 Ar and its special probes offers fundamental advantages over conventional coagulation methods.

 

The Argon Plasma beam permits non-contact coagulation of tissue, homeostasis of large surface bleeding can be stopped effectively and safely. And penetration depth is lesser than 3mm, no lead to carbonization.

 

Features:

A broad spectrum of applicators in open surgery and for endoscopy

Non-contact application, with no adhesion between applicator and tissue, and good visibility of the coagulation process.

Uniformly effective, wide-area hemostasis and coagulation.

Limited penetration depth, so that application in areas at risk of perforation is superior to all other coagulation methods.


Application:

It is applicable to ESD, EMR, ERCP, POEM, polypectomy, polyp burning, gynecology, orthopedics and surgery.

Open surgery

General surgery: hemostasis of large surface bleeding

Liver and gut surgery: liver transplantation,etc.

Cardiac and thoracic surgery: coronary artery bypass graft surgery, etc.

Traumatology and orthopedics: hemostasis of haemangioma

Parenchyma and bonetumor surgery: devitalization of cancer cell

 

Endoscopic operation

Gastrointestinal diseases, including tumorous stenosis and ulcer of the esophagus (i.e. oesophageal carcinoma), angiodysplasia and gastrointestinal bleeding (i.e. polypus, gastritis verrucosa, adenoma, malignant gastrointestinal tumors), palliative treatment of end-stage colorectal carcinoma respiratory disease, such as hemoptysis and neoplastic airway obstruction.

Laparoscopic general surgery: Laparoscopic liver resection, laparoscopic cholecystectomy, etc.

Gynaecologic laparoscopic surgery: hemostasis of large surface bleeding and devitalization of cancer cell

Endoscopic otolaryngology: endoscopic hemostasis and devitalization of cancer cell.

 

Character:

Both hand control and foot control outputs are available.

This machine has a power-off protection circuit, which can memorize the output set value of each function in real time.

Full-function automatic power compensation.

The high-frequency generator has a universal endoscope interface, which is used with gastroscope, enteroscope, laparoscope and other endoscopes.

Adopt ENDO CUT alternative output endoscope mode to provide safe and optimal coagulation effect under endoscope.

Output 10 groups of memory modes, support endoscopic treatment, ESD, ERCP, polypectomy and other operations.

With adjustable high-frequency pulse cutter system, the damage to tissues is minimized, and smokeless cutting.

When the air tube or hand piece tube was blocked or twisted, an alarm indication will be sent and the argon output will be cut off.

When the abdominal pressure pipe was blocked or twisted, an alarm indication will be sent and the argon output will be cut off.

Triple foot switch is controlled independently, convenient and fast.

The instrument is equipped with an argon pressure gauge to accurately and real-time display the argon working pressure.

Passed EMC test.

 

 Technical data:

 

Output Flow Indication 0.1L/min ~ 5L/min
(Optional 0.1L/min ~ 10 L/min or Max. 12 L/min)
Step Distance of Flow Control Varied step distance choices:
0.1L/min (0.1~2.0L/min);
0.5L/min (2.0-5.0L/min);
Security Check Power on self-test of hardware units and alarm system
Working Gas 99.99% purity medical argon
Argon Supply Support direct connection to the 40L type argon cylinder from hospital
Cylinder Capacity 4L
Scope Pressure Monitoring An environment pressure monitoring system, which automatically stops the match and gives an alarm once the pressure is above-threshold in endoscopy
Friendly Interface for Instruments Each Instruments with its own independent interface to prevent misconnection
Argon Independent output control Function Only under “spray coagulation” mode and press argon output switch, can it start to work; It prevents the electrode from being burned and the normal tissue from being hurt in case of miss operation of the switch in gastrointestinal endoscopy
ENDO CUT Function It combines the APC and Electrosurgery, which cut polypus and coagulate bleeding at the same time.
Type CF
Security Level I
Input voltage 220v~240v
Frequency of Electric Power 50/60Hz

 

Rated output power:

Pure cut:1W~300W (load 500 Ω);
A high-frequency alternating current yields smooth, rapid cuts that evoke little to know hemostasis.

Blend cut 1:1W~200W (load 500 Ω);
Preset blended cut modes give the surgeon varying degrees of hemostasis in cut mode.

Blend cut 2: 1W~150W (load 500 Ω);
The mode is used for vaporization of prostatic tissue or any fat tissue. The higher output wattage gives fast cutting effect of the tissue.

ENDO CUT
The mode is used to automatically controlled cut system to reduce the complication rate of endoscopic sphincterotomy (EST) and serum hyperamylasemia after EST compared to the conventional blended cut mode.

Spray coagulation:1W~80W (load 500 Ω);
Spray output is made for direct coagulation at the tissue with spark. above spark does not give any cutting effect as its CREST FACTOR is high.

DESICATE coagulation:1W~120W (load 500 Ω);
Coagulation in which the active electrode is in direct contact with the tissue is referred to as desiccation and is the type of coagulation used in most surgeries it ensures pinpoint desiccation with less destruction of peripheral tissue.

Bipolar coagulation (Optional):1W~70W (load 100 Ω);
For highly precise control over coagulation process and privations of carbonization.

Argon:
Dual Argon airflow adjustment, accuracy: 0.1 L/min; Argon pressure range: 0.20 ~ 0.50Mpa, Automatic compensation function stabilize output power, make depletion to its lowest.

Operation:

Ambient temperature range: 5 °C ~ 40 °C;

Relative humidity range: ≤ 80%;

Atmospheric pressure range: 86.0 ~ 106.0kPa;

Power supply: 110V ~ 220V ± 10%

Operating frequency: 416 kHz;

Power supply: 5A, 220V~±10%, 50±1% Hz;

Power consumption: ≤1100VA. (Cut function 300W);

 

Transportation and storage:

Ambient temperature range: -40°C~ 55°C;

Relative humidity range: ≤95%;

Atmospheric pressure range: 50.0~106.0kPa.

 

Configuration list

Main unit 1 PC□
Disposable neutral plate 5PCS□
Disposable neutral plate cable 1 PC□
Argon electrode 2 PC□
Triple foot switch 1 PC□
Power cable 1 PC□
High-frequency cable 1 PC□
Argon regulator 1 PC□
Argon hose 1 PC□
Argon bottle 1 PC□
Trolley 1 PC□
GBXP5*20(5A) Fuse 2PCS□
PACKING LIST 1 PC□
USER’S MANUAL 1 PC□
WARRANTY CARD 1 PC□
QUALITY CARD 1 PC□

 

 

Related Products

SMT-050DH Double column Digital Radiography System (Hand-driven)

Double column Digital Radiography System adopts large span independent ball tube frame design, easy installation, small occupation area, convenient maintenance. Four-way floating bed surface, easy to use and positioning. The lifting and lifting of X-ray source and bidirectional follow-up of horizontal position of bed column can reduce the positioning time and meet the requirements of large-flow clinical examination. Applicable to orthopedics, emergency room, operating room, etc.,As human head, limbs, chest, spine, lumbar spine for medical diagnosis

VSM-3200T 32kW 400mA Veterinary Digital Radiography System

1. Sophisticated technology, high rotating speed, good heat dissipation, high heat capacity, suitable for long-time clinical examination, long product life. 2. High frequency generator with high X-ray inverse frequency, large power and good X-ray quality. 3. Intelligence high voltage control system with various APR. 4. New generation a-Si Csl flat panel detector with high conversion rate achieves digital image acquisition with low dose. 5. High standard digital image processing system provides more post-processing and meets clinical needs.

SM-800L 2-8°C Medicine Refrigerator

Meet the requirements of the new version GSP for medicine storage. Can be used for medicine, reagents and various items that need cold storage. Suitable for hospitals, clinics, blood stations of health and epidemic prevention systems, university laboratories, etc.

SMH-60B 5-part Auto Hematology Analyzer

1. Throughput 60T/H.       2. 14-inch touch screen. 3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting. 4. 3D holographic scattergram displays the accurate 5-part differentiation of WBC. 5. Large storage capacity: 300,000 results(including histogram, scattergram, patient information).

Leave A Message


    Products
    Contact
    WhatsApp
    Email