Technical Specification
| No. | Item | Parameter Description |
| 1. | Model | SMB-50 |
| 2. | Power | 6kVA (Allowable deviation ±1%) |
| 3. | Inner Chamber Dimensions (mm) | 600*350*240 (Allowable deviation ±8%) |
| 4. | External Dimensions (mm) | 730*690*500 (Allowable deviation ±5%) |
| 5. | Main Control System | Controlled by a renowned brand PLC for more precise temperature control with smaller deviation; stable operation, simple and easy-to-understand operation. |
| 6. | Display Screen | High-resolution 4.3-inch color touchscreen allows direct on-screen operation; simple and aesthetically pleasing. |
| 7. | Inner Chamber Material | High-quality 304 stainless steel, constructed via argon arc welding with full-plate welding. Durable with a service life of 8 years. |
| 8. | External Cabinet Material | High-quality galvanized sheet with baked paint finish; smooth paint surface, uniform color, resistant to chipping). |
| 9. | Heating Principle | Uses carbon fiber far-infrared heating tubes as the heat source, enabling the sterilization chamber to rapidly reach the preset temperature with fast heating speed and high temperature control accuracy. Short sterilization time, no residue on sterilized items. |
| 10. | Operating Conditions | Ambient Temperature: 5°C to +40°C; Relative Humidity: Not exceeding 80%. |
| 11. | Sterilization Time | Within 160°C-209°C: 19-29 minutes (depending on material); Within 210°C-300°C: 3-19 minutes (depending on material). |
| 12. | Printing Function | Built-in mini printer for printing sterilization records, including physical curve printing function. Printed records can be stored for over 5 years. |
| 13. | Alarm System | Directly displays alarm information when a fault occurs, viewable in the alarm log. Over-temperature protection system: automatically cuts off power and stops operation when exceeding 310°C. |
| 14. | Cooling System | Automatically initiates cooling program after sterilization completes. The door can be opened and sterilized items can be removed using the provided clamp once cooled below 40°C. |
| 15. | Filtration System | Filter efficiency for particles with a diameter ≥0.3μm is not less than 99.5%. |
| 16. | Post-Use Maintenance | Inner and outer surfaces can be wiped with a soft cotton cloth (or sponge) using a neutral cleanser. |
| 17. | Scope of Application | Suitable for laboratory glassware, metal instruments and tools from operating rooms, dental departments, ophthalmology departments, supply rooms, and other products that can withstand high temperatures. |
| 18. | Power Supply | 220V/50Hz |