• Parameters

Technical Specification

Power requirement: AC220V-230V/50Hz or AC110V-120V, 50/60Hz, 700VA

Auxiliary power output: AC220V-230V/50Hz, max current 0.3A

 

Radiant Warmer Module

Skin temperature control range: 34.5℃-37.5℃

Skin temperature display range: 5°C-65°C

Deviation between the measure baby temperature by the sensor and control temperature: ≤0.5℃

Accuracy of skin temperature sensor: ±0.2℃

Temperature uniformity of mattress: ≤2℃

Mattress size: 735mm×590 mm

Failure alarm: power-off alarm, over-temperature alarm, sensor alarm, deviation alarm, setting alarm, check alarm and so on

Working noise: ≤50dB(A)

APGAR timer: audible and visible alarm when the device runs to 50″~1′, 4′50″~5′, 9′50″~10′

Angle of heater head: 0°, 30°, 60°for two ways

Inclination of the bassinet: ±12°

Environmental Requirement

Operating range: 18°C-30°C

Relative humidity: 30%RH-75%RH

Atmospheric pressure range: 800hPa-1060hPa

Ambient air flow speed: < 0.3m/s

 

Product Dimension

Dimension:

L1100mm × W730mm × H1795 mm (with stand)

L1100mm × W730mm × H1795 mm-1935mm (with VHA stand)

Height from the bassinet to the floor: 890mm

Wheel size(diameter): 150mm

Panel size:

left and right: 746mm × 175mm

front and rear: 580mm × 175mm

Weight:

105kg (with stand)

140kg (with VHA Stand)

 

Function Introduction

Three control Modes: Pre-warm mode, manual mode and baby mode control by micro-computer

Set temperature and skin temperature can be displayed separately

Horizontal angle of warmer module and inclination of the bassinet can be adjusted

The panels surrounded bassinet can be turned outward or disassembled

The height of whole machine can be adjusted from both sides to meet the operations of medical staff with different heights

Skin temperature sensor shedding alarm prompt function

Data storage function

X-ray cassette under the bassinet

APGAR timer

Multiple failure alarm indications

Related Products

SMT-050DH Double column Digital Radiography System (Hand-driven)

Double column Digital Radiography System adopts large span independent ball tube frame design, easy installation, small occupation area, convenient maintenance. Four-way floating bed surface, easy to use and positioning. The lifting and lifting of X-ray source and bidirectional follow-up of horizontal position of bed column can reduce the positioning time and meet the requirements of large-flow clinical examination. Applicable to orthopedics, emergency room, operating room, etc.,As human head, limbs, chest, spine, lumbar spine for medical diagnosis

VSM-3200T 32kW 400mA Veterinary Digital Radiography System

1. Sophisticated technology, high rotating speed, good heat dissipation, high heat capacity, suitable for long-time clinical examination, long product life. 2. High frequency generator with high X-ray inverse frequency, large power and good X-ray quality. 3. Intelligence high voltage control system with various APR. 4. New generation a-Si Csl flat panel detector with high conversion rate achieves digital image acquisition with low dose. 5. High standard digital image processing system provides more post-processing and meets clinical needs.

SM-800L 2-8°C Medicine Refrigerator

Meet the requirements of the new version GSP for medicine storage. Can be used for medicine, reagents and various items that need cold storage. Suitable for hospitals, clinics, blood stations of health and epidemic prevention systems, university laboratories, etc.

SMH-60B 5-part Auto Hematology Analyzer

1. Throughput 60T/H.       2. 14-inch touch screen. 3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting. 4. 3D holographic scattergram displays the accurate 5-part differentiation of WBC. 5. Large storage capacity: 300,000 results(including histogram, scattergram, patient information).

Leave A Message


    Products
    Contact
    WhatsApp
    Email