Technical Specification
Lead: Standard 12 leads
Lead acquisition: 12 leads synchronously acquisition
Work mode: Manual/Auto/R-R/Store
Filter:
AC Filter: 50Hz/60Hz
EMG Filter: 25Hz/45Hz
Anti-Drift Filter: 0.15Hz(adaptive)
CMRR: ≥120dB(with AC Filter )
Input Circuit: Floating; Protection circuit against defibrillator effect
Input impedance: ≥50MΩ
Input circuit current: ≤0.05ųA
Patient current leakage: <10ųA
Calibrating voltage: 1mV±2%
Voltage tolerance: ≥±500mV
Time constant: ≥3.2s
Frequency response: 0.05~160Hz(-3db)
Noise level: <15ųVp-p
Inter-channel interference: ≤0.5mm
Sensitivity: Auto,2.5, 5, 10, 20, 40mm/mV(±3%)
Recording mode: 3ch+++, 6ch, 6ch+, 12ch
Paper speed: 6.25, 12.5, 25, 50mm/s(±3%)
Recording Paper: 210mmx30m roll thermal paper
Power supply
AC:110-230V(±10%),50/60Hz,40VA
DC: Built-in rechargeable Li-ion battery, 14.4V, 4400mAh
Communication Port: USB, UART
Dimension/Weight: 347mmx293mmx83mm, 4.8kgs
Packing List
Main unit 1set
Patient cable 1pc
Limb electrodes 1set ( 4pcs )
Chest electrodes 1set ( 6pcs )
Recording paper 1roll ( 210mmx30m)
Fuse 2pcs
Paper axis 3pc
Grounding cable 1pc
Power cable 1pc
Operation manual 1copy